Measurements
R&S
®
RTO
125User Manual 1316.0827.02 ─ 06
5. Select "Show labels".
6. To display only the crossing points of the cursors with the waveform, select the cursor
style "Rhombus".
To display both the crossing points and the cursor lines, select the cursor style "Line
& Rhombus".
5.1.3 Reference for Cursor Measurements
Cursor measurements are configured in the "Cursors" dialog box which is opened via the
"Cursor > Setup" menu or the "Cursor Results" box, or by pressing the CURSOR key.
5.1.3.1 Cursor Setup Tab
This tab contains general settings for cursor measurements.
C1/C2/C3/C4
The settings for each of the four available cursor measurements are configured on sep-
arate tabs. For each measurement, a horizontal pair of cursors, a vertical pair of cursors,
or both can be displayed.
Enable
Enables the selected cursor measurement.
Source
Defines the source of the cursor measurement. Any of the input signal, math, reference
or XY waveforms can be selected.
SCPI command:
CURSor<m>:SOURce on page 517
Type
Defines the cursor type to be used for the measurement.
Cursor measurements