Measurements
R&S
®
RTO
156User Manual 1316.0827.02 ─ 06
Upper level
The upper reference level, required e.g. to determine a rise - for manual reference level
mode, absolute level definition and user reference level.
SCPI command:
REFLevel<m>:ABSolute:ULEVel on page 534
MEASurement<m>:REFLevel:RESult:UPPer on page 540
Lower level
The lower reference level, required e.g. to determine a fall - for manual reference level
mode, absolute level definition and user reference level.
SCPI command:
REFLevel<m>:ABSolute:LLEVel on page 535
MEASurement<m>:REFLevel:RESult:LOWer on page 540
Histogram averaging
Enables averaging over several histograms to determine the reference levels.
This function is only available in automatic reference level mode.
SCPI command:
REFLevel<m>:AUTO[:STATe] on page 531
Average Count
Defines the number of histograms to calculate the average from.
This function is only available in automatic reference level mode.
SCPI command:
REFLevel<m>:AUTO:COUNt on page 532
Hysteresis Tab
This tab allows you to define a hysteresis for measurements that determine zero-cross-
ings.
Automatic Measurements