Remote Control
R&S
®
RTO
445User Manual 1316.0827.02 ─ 06
PROBe<m>:SETup:OFFSet:AZERo
Performs an automatic correction of the zero error. If the DUT is ground-referenced, the
Auto Zero function can improve the measurement results.
See also: "Auto Zero, AutoZero DC offset, Use AutoZero" on page 46
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
PROBe<m>:
SETup:OFFSet:TOMean
Performs an automatic compensation for a DC component of the specified input signal
using the result of a background mean measurement.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
PROBe<m>:SETup:MODE <Mode>
Select the action that is started with the micro button on the probe head.
See also: "Micro button action" on page 46.
Suffix:
<m>
.
1..4
Selects the input channel.
Command Reference